Test of two-dimensional position sensitive detector
A two-dimensional position sensitive detector (PSD) which will be used in the measurement of the radioactive ion beam has been tested. By carefully selecting the pulse shaping time contants in the electronics and improving the method of data analysis; i.e. as the main amplifier’s shaping time was more than 6μs and position signal’s shaping time was 0.5—1μs
and used the energy signal as the denominator in the formula of position
better position resolution
energy resolution and position linearity have been obtained. The result shows that these type detectors can be used in the measurement of low energy radioactive ion beam.