SUN Chaoming, TANG Guangping, LI Qiang, et al. Comparative analysis and measurement of focus spot for X-ray source[J]. Nuclear techniques, 2015, 38(11): 110202
SUN Chaoming, TANG Guangping, LI Qiang, et al. Comparative analysis and measurement of focus spot for X-ray source[J]. Nuclear techniques, 2015, 38(11): 110202 DOI: 10.11889/j.0253-3219.2015.hjs.38.110202.
射线源焦点的大小与形状对射线检测图像质量有重要的影响,但在射线检测中却很少对射线源焦点进行实测,这不利于准确控制射线检测图像质量。对射线源焦点的测量方法进行对比试验分析,以获得对射线源焦点进行评估的有效便捷方法。分别应用小孔成像板、星形测试卡、双线型像质计(Image quality indicator
IQI)等测试体,对工业应用的便携式X射线机进行了焦点尺寸测量。用星形测试卡、双线型像质计所测算的焦点尺寸一致性较好,但较之小孔成像的测量结果却存在较大偏差。实际检测时射线机焦点大小可能与标称值明显不同,有必要对射线焦点特性进行及时评价;双线型像质计可一定程度上代替星形测试卡的使用,便捷地实现图像不清晰度的测量、射线源焦点尺寸的评估。
Background: Image quality in radiographic testing is often conditioned by the size and distribution of X-ray source. However
determination of focal spot is not performed effectively in practice
thus it is difficult to accurately control the image quality in the testing. Purpose: This study aims to obtain a quick and efficient test method to evaluate the focal spot for X-ray source. Methods: Test tools such as pinhole plate
star test pattern and duplex wire image quality indicator (IQI) were used to measure the focal spot of a portable X-ray source separately. Several measurements were carried out and compared in details. Results: Results of star test pattern and duplex wire IQI were coincident
but different from the results of pinhole imaging method obviously. Conclusion: Nominal focal spot size of an X-ray source may be considerably different from that in practice use
so it is necessary to timely evaluate characteristics of the focal spot. The duplex wire IQI can usually replace star test pattern to measure un-sharpness in radiographic testing images and to evaluate focal spot size. Abstract