利用多能量γ射线源
152
Eu和
133
Ba测得高纯锗探测器在20 cm处的效率刻度曲线
结合单能量γ射线源
241
Am和
137
Cs获得的距离递推关系
得到高纯锗探测器在2 cm处无符合效应的效率刻度曲线。与MCNP模拟计算结果比较
获得
60
Co在2 cm处的符合相加修正因子。
By using multi-gamma ray sources of
Eu and
Ba placed 20 cm from the HPGe detector
its detection efficiency was obtained.Combined with the distance recurrence relation
which was derived by single-gamma ray sources of
Am and
Cs
the detector efficiency calibration curve at 2 cm could be deduced.Finally
by comparing the experiment results with the simulation results by the MCNP code
the coincidence factor could be determined.