ZHANG Lu1 HE Chaohui1 CHEN Liang2 ZHANG Zhongbing2 LIU Jinliang2 REN Xiaotang3 1. In-site measurement of intensity of proton beam using film scintillator[J]. Nuclear techniques, 2012, 35(1): 21-25.
ZHANG Lu1 HE Chaohui1 CHEN Liang2 ZHANG Zhongbing2 LIU Jinliang2 REN Xiaotang3 1. In-site measurement of intensity of proton beam using film scintillator[J]. Nuclear techniques, 2012, 35(1): 21-25.DOI:
precise measurement of proton beam current is important for reliability and accuracy of the experiments.Traditional detection methods
such as Faraday cup
Si(Au) surface barrier detector
and diamond detector
are of beam-block type measurements
where changes in proton beam current can hardly be monitored in-site.In this paper
film scintillators are used to detect the proton beam.Protons travel through the film scintillator
deposit part of the energy and make the scintillator emit lights
which are collected by a photomultiplier tube.Rresponse of the film scintillator to protons are calculated and verified by experiments with proton beams at 3–10 MeV from the 2×6 MeV tandem accelerator at Peking University.