the radon-thoron discriminative detector improved by our laboratory was calibrated at Radon Laboratory of University of South China
and participated in the international in- tercomparison organized by NIRS(National Institute of Radiological Sciences)in Japan. The calibration result demon- strated that correlation between the exposure level and the track density was fairly good for both radon and thoron
with the correlation coefficient of 0.9894 and 0.9885
and the calibration factors of 4.27 ± 0.40 and 2.33 ± 0.39 tracks·cm–2·(kBq·m–3·h)–1 for radon and thoron
respectively. According to the results of inter-comparison
in pure ra- don
relative percent difference from the reference value (RPD) was 12% and coefficient of variation (COV) was 7.7%
ranking category B of NRPB
and was in the control level of EPA; while in high 220Rn condition
RPD and COV for thoron were 22% and 4.8%
respectively. Concerning the higher deviation for thoron measurement