QI Huirong1, 2 WANG Jinchuan1 XIAO Guoqing1 GUO Zhongyan1 ZHAN Wenlong1 XU hushan1 LI Jiaxing1 SUN Zhiyu1 HU Zhengguo1 MAO Ruishi1, 2 LI Chen1, et al. Study of timing voltage divider system of PMT for neutron wall at CSR terminal[J]. Nuclear techniques, 2005, (6): 472-475.
QI Huirong1, 2 WANG Jinchuan1 XIAO Guoqing1 GUO Zhongyan1 ZHAN Wenlong1 XU hushan1 LI Jiaxing1 SUN Zhiyu1 HU Zhengguo1 MAO Ruishi1, 2 LI Chen1, et al. Study of timing voltage divider system of PMT for neutron wall at CSR terminal[J]. Nuclear techniques, 2005, (6): 472-475.DOI:
On the outer-target experimental terminal in Heavy Ion Cooling Storage Ring at Lanzhou
an experimental setup for the study of RIB physics and the investigation of features of the asymmetric nuclear matter by means of RIB is highly required. The neutron detection wall is a main part of the detection equipment
in which there are 1000 photomultipliers serving as signal readout devices. The PMTs’ voltage divider is studied. Both design parameters and the measured results show that the timing-type voltage divider is more suitable for detecting time signals.